Sample excerpt does not include step-by-step text instructions for performing this procedure
The full content of this section includes:
- Step-by-step text instructions for performing the procedure
- Clinical pearls providing practical clinical tips from medical experts
- Patient safety guidelines consistent with Joint Commission and OHSA standards
- Links to medical evidence and related procedures
|
Figure 6
:
Placement of a double-lumen endotracheal tube. A, Insert the tube with the endobronchial lumen pointing forward (up). B, Gently advance, rotating the tube to the appropriate side. C, Inflate the endobronchial and endotracheal cuffs and test for proper position. From Miller RD (ed): Miller’s Anesthesia. Philadelphia, Elsevier, 2005.
Figure 7
:
Functional test for double-lumen endotracheal tube position. With the tube positioned in too far on the left side and both cuffs inflated, breath sounds will be heard on the left when the right side lumen is clamped, will be diminished or absent on both sides with the left side lumen clamped, and will be heard on the left with the left side lumen clamped and the endobronchial cuff deflated. With the tube out too far and both cuffs inflated, breath sounds will be heard on the left and right with the right-side lumen clamped, will be diminished or absent on both sides with the left-side lumen clamped, and will be heard on both sides with the left-side lumen clamped and the endobronchial cuff deflated. With the tube positioned out too far into the right side and both cuffs inflated, breath sounds will be heard on the right with the right-side lumen clamped, will be absent or diminished on both sides with the left-side lumen clamped, and will be heard on the right with the left-side lumen clamped and the endobronchial cuff deflated. From Miller Ronald D (ed): Miller’s Anesthesia. Philadelphia, Elsevier, 2005.
Figure 8
:
Herniation of the endobronchial cuff across the carina; this happens with overinflation of the endobronchial cuff and/or malposition of the double-lumen tube with the endobronchial cuff too near the carina. From Miller RD (ed): Miller’s Anesthesia. Philadelphia, Elsevier, 2005.
Figure 9
:
Bubble test. Good seal, no bubbles seen. From Miller RD (ed): Miller’s Anesthesia. Philadelphia, Elsevier, 2005.
Figure 10
:
Bubble test. Poor seal, bubbles seen. Adjustment of endobronchial cuff inflation or repositioning of the double-lumen tube is indicated. From Miller RD (ed): Miller’s Anesthesia. Philadelphia, Elsevier, 2005.
|