Sample excerpt does not include step-by-step text instructions for performing this procedure
The full content of this section includes:
- Step-by-step text instructions for performing the procedure
- Clinical pearls providing practical clinical tips from medical experts
- Patient safety guidelines consistent with Joint Commission and OHSA standards
- Links to medical evidence and related procedures
|
Figure 14
:
Anesthetize the skin with a small amount of anesthetic.
Figure 15
:
After assembling the needle, diaphragm chamber, and syringe, place it into the Pressure Monitor.
Figure 16
:
Hold the device at the intended angle of insertion, and press the “zero” button.
Figure 17
:
Proper needle position for the anterior compartment.
Figure 18
:
Proper needle position for the deep posterior compartment.
Figure 19
:
Proper needle position for the lateral compartment.
Figure 20
:
Proper needle position for the superficial posterior compartment.
Figure 21
:
Proper needle position for the volar compartment.
Figure 22
:
Proper needle position for the dorsal compartment.
Figure 23
:
Proper needle position for the mobile wad compartment.
Figure 24
:
For the gluteal compartments, insert the needle in the region of maximal tenderness.
Figure 25
:
Proper needle position for the medial compartment.
Figure 26
:
Proper needle position for the central compartment. Note that the entry site is the same as for the medial compartment; however, the needle is advanced to a depth of 3 cm.
Figure 27
:
Proper needle position for the lateral compartment.
Figure 28
:
Proper needle position for the intrinsic compartments.
|